Comparison of axial length measurements

Difference* Difference (Lenstar® standard-5 LS – IOLMaster® composite-5 IM) Difference (Lenstar® standard-5 LS – IOLMaster® composite-20 IM) Difference (IOLMaster®composite-5 IM – IOLMaster® composite-20 IM)
AL (mm)
Mean ± SD 0.03±0.03 0.02±0.03 −0.0±0.03
Range −0.06 to 0.19 −0.08 to 0.15 −0.20 to 0.08
P-value paired-t comparing means <0.001 <0.001 0.840

Notes:

*

Based on eyes with measurements from both methods. Lenstar® standard-5, Haag-Streit (Koeniz, Switzerland); IOLMaster® composite-5; IOLMaster® composite-20, Carl Zeiss Meditec AG (Jena, Germany).