Table 2: A summary of impedance comparisons between this study and others.

Reference Electrode Type Mammal and insertion location Contact surface area (mm2) 1 kHz impedance (ohms) Impedance/unit surface area (ohms/mm2)
Williams et al. (2007) Microelectrode tungsten array Hooded Long Evan Rat: S1 cortical barrel 0.112–0.143 ≈100,000 698,000 (if maximum contact surface area)
Lempkaet al. (2009) custom Pl/Ir microelectrode Rhesus macaque: ventral thalamus and subthalamic nucleus 1.18 ≈3500 2971
This work Medtronic model 3387 Agarose 5.99 (@ 1.172 kHz) 3186.27 (bipolar) 532.4