Potential failure modes |
Simulation |
Incorrect/inadequate immobilization, patient preparation or documentation |
Faulty image data |
Data transfer (images and other DICOM data) |
Errors in data transfer between imaging systems and TPS (human, software, hardware) |
Treatment planning directive (MD orders) |
Incorrect/inadequate specification of image sets, protocols, registration goals, guidelines, dose limits, or fractionation for planning or delineation |
Incorrect/inadequate documentation of previous treatments |
TP anatomy |
Errors in image import and registration |
Incorrect/inadequate critical and target structure delineation and construction |
Inaccurate density map/correction |
Errors in creating, using, editing, or saving anatomical models |
Incorrect/inadequate dose point definitions |
Treatment planning |
Incorrect/inappropriate region of interest (ROI) volumes for optimization |
Incorrect/incomplete/inappropriate prescription, objectives, planning/treatment techniques, beam energy, machine, isocenter, or dose calculation parameters |
Errors in optimization, dose calculation, heterogeneity correction, leaf sequencing, or plan evaluation |
Plan approval and preparation |
Bad or incorrect plan approved |
Incorrect/inadequate localization images and protocols planned |
Errors in plan information indicated for treatment |
Treatment |
Patient medical condition changes between prescription and treatment |
Incorrect patient, records, plan data, isocenter, position, treatment parameters/accessories |
Errors in setup, immobilization, localization, treatment parameter data, machine operation, or treatment delivery |