Table 2.
Magnetic field homogenization of a central 12 mm thick slab of the in vivo rat brain at 11.7 Tesla with different shimming techniques. Theoretical shimming including various spherical harmonic (SH) orders was applied statically and dynamically assuming unlimited amplitude range. DYNAMITE shimming was based on experimentally calibrated basis fields with the presented MC-RF setup including a ±1 A current limitation on all MC channels. The average standard deviation (mean ± SD) of the magnetic field distributions in 12 rat brains after shimming is reported.
uncorrected | 136 ± 20 Hz |
---|---|
| |
static SH 1st | 69 ± 8 Hz |
static SH 2nd | 61 ± 11 Hz |
static SH 3rd | 52 ± 7 Hz |
static SH 4th | 44 ± 8 Hz |
static SH 5th | 38 ± 7 Hz |
| |
SH-DSU 1st | 42 ± 4 Hz |
SH-DSU 2nd | 33 ± 4 Hz |
SH-DSU 3rd | 27 ± 3 Hz |
SH-DSU 4th | 22 ± 3 Hz |
| |
static MC shim | 40 ± 5 Hz |
DYNAMITE shim | 17 ± 3 Hz |