Skip to main content
. 2014 Jul 30;7:263–271. doi: 10.2147/MDER.S66848

Table 1.

Burst pressure medians for small and large diameter porcine carotid arteries with comparisons between Harmonic ACE®+7 and LigaSure™ (comparisons were performed with a Mann–Whitney test)

Vessel diameter LigaSure Harmonic 7 power level 3 Harmonic 7 advanced hemostasis
3–5 mm 836 mmHg 1,078 mmHg
P=0.046
1,314 mmHg
P=0.001
5–7 mm 591 mmHg NA 1,419 mmHg
P<0.001

Notes: Harmonic ACE+7 (Ethicon Inc., Cincinnati, OH, USA). Ligasure (Covidien, Mansfield, MA, USA).