Table 1.

Anatomical criteria from the instructions for use for AAA endovascular devices approved by the FDA

Endovascular device Year
of FDA
approval
Neck
diameter
(mm)
Neck
length
(mm)
Neck
angulation
(°)
Iliac neck
length
(mm)
Iliac neck
diameter
(mm)
Ancure™ (EndoVascular Technologies, Inc., USA)* 1999 18–26 ≥15 NS ≥20 <13.5
AneuRx® (Medtronic Vascular, Inc., USA) 1999 18–25 ≥10 ≤45 NS NS
Excluder® (W. L. Gore & Associates, Inc., USA) 2002 19–26 ≥15 ≤60 ≥10 10–18.5
Zenith® (Cook Medical Technologies, USA) 2003 18–28 ≥15 ≤60 ≥15 10–20
Low-permeability Excluder® (W. L. Gore & Associates, Inc., USA) 2004 19–26 ≥15 ≤60 ≥10 10–18.5
Powerlink® (Endologix, Inc., USA) 2004 18–26 ≥15 ≤60 ≥15 8–18
Enlarged-neck Zenith® (Cook Medical Technologies, USA) 2006 18–32 ≥15 ≤60 ≥15 10–20
Talent® (Medtronic Vascular, Inc., USA) 2008 18–32 ≥10 ≤60 ≥15 8–22
Enlarged-neck Powerlink® (Endologix Inc., USA) 2009 18–32 ≥15 ≤60 ≥15 10–23
Enlarged-neck Excluder® (W. L. Gore & Associates, Inc., USA) 2009 19–29 ≥15 ≤60 ≥10 10–18.5
Endurant® (Medtronic Vascular, Inc., USA) 2010 19–32 ≥10 ≤60 ≥15 8–25
Ovation® (Trivascular, Inc, USA) 2011 15.5–30 ≥7 45–60 ≥10 8–20
Fenestrated Zenith® (Cook Medical Technologies, USA) 2012 19–31 ≥4 <45 >30 7–21
Aorfx® (Lombard Medical, UK) 2013 19–29 ≥15 ≤90 ≥15 9–19
*

Device discontinued in 2003.

Changed to ≥15 mm in 2003.

Abbreviations: AAA, abdominal aortic aneurysm; NS, not specified.