Published in final edited form as: J Biomech. 2014 Jun 17;47(11):2584–2589. doi: 10.1016/j.jbiomech.2014.05.027

Table 1.

The structural properties of Restorelle (n=5) and UltraPro (n=4) subjected to a uniaxial tensile testing protocol. Mesh samples were loaded to failure along two perpendicular directions orientation 1 (O1) and orientation 2 (O2), and compared utilizing a Student's t-test (p-value<0.05). O1 represents the direction prolapse meshes are typically implanted.

Mesh Failure Load (N) Relative Elongation (%) Low Stiffness (N/mm) High Stiffness (N/mm) Inflection Point (%)
Restorelle Orientation 1 22.7 ± 2.82 68.4 ± 6.34 0.177 ± 0.033 0.582 ± 0.082 29.1 ± 2.1
Restorelle Orientation 2 24.3 ± 3.32 68.4 ± 7.12 0.198 ± 0.038 0.613 ± 0.051 28.8 ± 2.2
P-value 0.44 0.99 0.37 0.49 0.80
UltraPro Orientation 1 7.83 ± 0.69 87.9 ± 5.62 0.009 ± 0.002 0.236 ± 0.019 46.5 ± 5.2
UltraPro Orientation 2 122.3 ± 13.8 54.4 ± 3.27 0.258 ± 0.085 4.75 ± 0.295 22.6 ± 1.86
P-value <0.001 <0.001 0.009 <0.001 <0.001