Skip to main content
. Author manuscript; available in PMC: 2014 Aug 14.
Published in final edited form as: Phys Med Biol. 2008 Jul 31;53(16):4455–4470. doi: 10.1088/0031-9155/53/16/016

Table 2.

Comparison of simulated PDDs to measured PDDs in the large snout. Ep is the proton energy at nozzle entrance, r90 is the nominal range at the distal 90% dose location, and m is the nominal SOBP width of the SOBP.

Ep
(MeV)
r90
(mm)
m
(mm)
Δr90
(mm)
Δm
(mm)
ΔDmax
(%)
250 250 40 0.04 0.09 2.2
250 250 100 0.40 0.75 1.6
250 250 160 0.42 5.44 2.0
250 207 100 0.90* 7.70*
225 206 40 1.00 1.10 3.0
225 206 100 0.50 1.50 2.0
225 206 140 0.80 3.00 1.7
225 166 100 1.50* 7.80*
200 165 160 2.17 0.01 1.5
200 165 100 2.15 2.46 1.9
200 165 140 2.27 1.97 1.8
200 138 60 1.60* 2.24*
180 137 40 2.05 1.89 2.3
180 137 60 2.12 0.59 2.0
180 137 100 2.22 1.35 2.5
180 111 60 1.13* 6.67*
160 110 20 2.05 0.17 3.0
160 110 40 2.03 2.76 2.8
160 110 60 2.02 3.47 3.0
160 110 80 2.07 2.13 2.2
160 110 100 2.11 0.23 1.0
160 85 60 1.06 3.47 3.0
140 84 40 0.61 0.78 1.9
140 74 60 0.69 1.30 2.9
120 63 40 1.07 1.85 1.9
120 63 20 1.24 0.12 2.6
100 43 30 1.17 0.04 2.3
100 35 20 1.28 1.02 2.5

Δr90 is the absolute difference between the simulated and the measured ranges at the distal 90% dose location, Δm is the absolute difference in SOBP width (Δr90 and Δm values with an ‘*’ were compared to the nominal values), and ΔDmax is the maximum absolute dose difference relative to the dose at the center of the SOBP.