Table 2. Optical performance characterization of a low-length profile MRFL prototype at 120mm working distance.
Wide-angle imaging probe | High-magnification imaging probe | |||
---|---|---|---|---|
Field | Center | Corner | Center | Corner |
Field of view (mm2) | 160 × 120 | 53 × 40 | ||
Limiting resolution (um) | 176.68 | 280.90 | 78.74 | 140.45 |
Limiting spatial frequency (lps/mm) | 2.83 | 1.78 | 6.35 | 3.56 |
Throughput (mm2sr) | 0.3351 | 0.0372 |