Table 2.
List of fast MRI pulse sequences that have been tested for CEST studies
Suitable Saturation schemes | Typical TR and TE | Example acquisition time per image (sec)a | Favorable B0 | |||
---|---|---|---|---|---|---|
Single long | Short repetitive | TR (ms) | TE (ms) | |||
RARE(4) | + | >5000 | ~50b | 24c | No limit d | |
FLASH(91) | + | 100-300 | <5 | 19.2e | No limit | |
EPI(51,132) | + | + | >5000 | ~50 | 40f | 1.5T-3T |
FISP(38) | + | 2 | ½ TR | 2.5g | 1.5T-3T | |
3D-GRASE(103) | + | 2500 | 200 | 20h | No limit |
using values in literature based on a 2D image with 64 phase encoding steps and a single average unless otherwise noted.
effective TE
with TR=6 sec and RARE factor =16.
except it may result in high SAR at high field
with TR=200ms and 32 dummy scans to reach steady state of saturation
four-shot EPI with TR=10 s
includes 10 dummy scans to reach steady state.
With SENSE acceleration factors of 2×2 in the RL (44 steps) and SI (28 steps) directions, SE factor = 22, EPI factor = 7.