TABLE I.

Ring “A”: Median and standard deviation of the three materials scanned using SAFIRE at decreasing doses. *P value < 0.05—Comparing median SAFIRE value and median WFBP value at the same dose level for the same material.

Material SAFIRE: Ring A—median HU (sd)
Dose (mGy) 11.94 9.04 5.97 3.04 1.55 0.74
Acrylic 120 (6.57) 120 (7.07) 121 (7.21) 120 (8.83) 121 (11.21) 120 (17.11)
Water −5 (6.53) −6 (6.82) −5.00 (7.71) −5 (9.43) −6 (11.33) −6 (15.97)
20 lb foam −702 (4.35) −703 (4.47) −702 (5.97) −703 (8.37) −703 (9.74) −703 (14.75)
12 lb foam −824 (3.13) −824 (3.56) −823 (4.10) −823 (5.92) −823 (7.93) −823 (11.67)
Lung foam −857 (6.82) −857 (6.97) −856 (7.53) −856 (8.97) −856 (11.30) −856 (15.43)
4 lb foam −937 (3.46) −938 (4.23) −938 (4.65) −938 (6.49) −937 (9.93) −937 (14.32)
Internal air −999 (2.97) −999 (3.30) −999 (4.09) −998.0 (5.23)* −997 (6.88)* −994 (9.08)*
External air −1002 (4.37) −1002 (4.93) −1002 (4.94) −1001 (6.54)* −1000 (8.01)* −996 (10.77)*