Skip to main content
. 2014 Aug 25;41(9):091910. doi: 10.1118/1.4893498

TABLE V.

Ring “C”: Median and standard deviation of the three materials scanned using SAFIRE at decreasing doses. *P value < 0.05—Comparing median SAFIRE value and median WFBP value at the same dose level for the same material.

Material SAFIRE: Ring C—median HU (sd)
Dose (mGy) 11.94 9.01 5.97 3.04 1.58 0.76
Acrylic 120 (6.60) 121 (6.89) 121 (7.32) 120 (10.11) 120 (12.60) 121 (18.47)
Water −8 (6.39) −7 (6.66) −6 (7.75) −8 (9.24) −7 (13.10) −8 (18.12)
20 lb foam −702 (4.15) −702 (4.38) −702 (6.12) −702 (7.69) −702 (11.70) −702 (16.71)
12 lb foam −824 (3.41) −824 (3.61) −823 (4.32) −824 (6.18) −823 (8.64) −823 (12.56)
Lung foam −857 (6.61) −856 (6.84) −856 (7.54) −856 (9.07) −856 (11.87) −856 (16.31)
4 lb foam −937 (4.04) −937 (4.65) −937 (5.09) −937 (7.85) −937 (9.92) −936 (15.02)*
Internal air −1000 (3.31) −999 (3.85) −999 (4.73) −998 (5.71)* −997 (7.54)* −993 (10.05)*
External air −1001 (4.63) −1001 (4.88) −1001 (5.31)* −1000 (7.31)* −998 (9.15) −994 (12.20)*