Skip to main content
. 2014 Aug 24;9:13. doi: 10.1186/1748-7161-9-13

Figure 3.

Figure 3

Evidence of screw loosening and proximal junctional kyphosis. CT of a 16 years old girl following segmental pedicle screw fixation T2-T12. Sagittal images (A) six weeks, and (B) two years following surgery of the same patient in Figure 2. Beside evidence of loosening, image B shows occurrence of proximal junctional kyphosis above the upper limit of the implant compared with image A.