Table 3.
Frequency of the occurrence of individual errors in N = 462 patients at t1 and t2, respectively, and the correlation to the different error classes
Possible sources of error | Error class | t1 | T2 | ||
---|---|---|---|---|---|
Number | Percent | Number | Percent | ||
[n] | [%] | [n] | [%] | ||
Squeezing out the blood | F1 | 216 | 46.8 | 77 | 16.7 |
Settings (date, time) | F5 | 209 | 45.2 | 65 | 14.1 |
Sideways pricking of the finger pad | F3 | 157 | 34.0 | 36 | 7.8 |
Hand-washing | F1 | 99 | 21.4 | 20 | 4.3 |
Inserting / changing the lancet (skill) | F3 | 83 | 18.0 | 22 | 4.8 |
Stimulating circulation (if needed) | F3 | 81 | 17.5 | 13 | 2.8 |
Drying the hands | F1 | 71 | 15.4 | 6 | 1.3 |
Adjusting the prick depth | F3 | 63 | 13.6 | 8 | 1.7 |
Closing the test strip container after taking one out | F2 | 59 | 12.8 | 15 | 3.3 |
Documentation / saving the result | F5 | 55 | 11.9 | 11 | 2.4 |
If disinfected, was the skin dry? | F1 | 52 | 11.3 | 10 | 2.2 |
Checking the coding | F1 | 41 | 8.9 | 8 | 1.7 |
Cleanliness of device and measuring cell | F2 | 37 | 8.0 | 14 | 3.0 |
Sufficiently large drop of blood extracted | F2 | 26 | 5.6 | 6 | 1.3 |
Applying the blood sample / absorption of the test field | F2 | 25 | 5.4 | 6 | 1.3 |
Quantity of the applied blood | F2 | 25 | 5.4 | 6 | 1.3 |
Changing the coding (if required) | F1 | 21 | 4.6 | 4 | 0.9 |
Loading the lancet | F4 | 18 | 3.9 | 3 | 0.7 |
Expiration date of the test strips | F1 | 15 | 3.3 | 5 | 1.1 |
Storage conditions of the test strips | F1 | 14 | 3.0 | 0 | 0 |
Inserting the test strip / inserting the disk | F4 | 14 | 3.0 | 5 | 1.1 |
Check with the glucose control solution | F1 | 13 | 2.8 | 3 | 0.7 |
Condition of the battery | F4 | 12 | 2.5 | 8 | 1.7 |
Basic handling of the measuring device | F4 | 4 | 0.9 | 0 | 0 |
Choice of the desired unit (mg/dL, mmol/L) | F2 | 2 | 0.4 | 2 | 0.4 |
Use of correct test strips | F1 | 0 | 0 | 0 | 0 |