Table II. Nano-LC-MS/MS instruments.

Mass spectrometer Q Exactive (Q-E) Orbitrap Elite Orbitrap Velos Pro
HPLC Thermo EASY-nLC 1000 Eksigent nanoLC Ultra-2D plus Thermo EASY-nLC 1000
Pre-column Acclaim® PepMap 100 pre-column 75 μm × 2 cm, 3 μm ReproSil-Pur 120 C18-AQ (Dr. Maisch GmbH, Germany)
Analytical column Acclaim® PepMap RSLC column 75 μm × 15 cm, 1.9 μm ReproSil-Pur 120 C18-AQ (Dr. Maisch GmbH, Germany) 75 μm × 15 cm, 3 μm ReproSil-Pur 120 C18-AQ (Dr. Maisch GmbH, Germany)
MS resolution 70,000 60,000 60,000
MS automatic gain control target 1e6 1e6 1e6
MS/MS HCD resolution 17,500 15,000 15,000
MS/MS automatic gain control target 1e5 HCD, 5e4; CID, 1e4 HCD, 5e4; CID, 3e4
MS/MS normalized collision energy 35.0 HCD, 37.5; CID, 40.0 HCD, 36.0; CID, 40.0

Listed are three instrument setups and settings for MS runs. For Elite and Velos, both high-energy C-trap dissociation (HCD) and collision-induced dissociation (CID) were used for MS/MS scans.