Table 1.

Between day precision and accuracy for calibrators and three different batches of QC samples. Data were collected from a replication experiment in which each sample was run in duplicate on a different day (n = 5–9). Concentrations of Aβ42 for QC samples lot#242 and 142 were obtained from the calibration curve prepared from standard material lot# 942

Calibrator (pg/mL) lot#942
Mean accuracy (%) Mean SD Mean CV (%) n
Nominal Reads
50 49.9 99.7 10.7 10.8 18
100 105.1 105.1 7.9 7.4 18
200 200.9 100.4 6.7 6.6 18
350 345.3 98.6 5.5 5.6 18
500 482.5 96.5 4.7 4.8 18
750 738.2 98.4 7.2 7.3 18
1000 1003.5 100.4 6.2 6.1 18
3000 3019.5 100.6 2.7 2.6 18
QC samples lot#942
250 255.6 102.2 24.9 9.7 14
400 404.7 101.2 33.3 8.2 14
800 793.6 99.2 50.7 6.4 14
QC samples lot#242
250 238.2 95.8 22.0 9.2 10
400 366.8 91.7 19.1 5.2 10
800 782.1 97.8 79.3 10.1 10
QC samples lot#142
250 249.0 99.6 24.2 9.7 10
400 402.4 100.6 29.8 7.4 10
800 770.4 96.3 62.7 8.1 10