Table 2.

Table 3. Bias of QuarkRMR® relative to Deltatrac® and within multiple measurements with Deltatrac® (95% confidence intervals of the differences that exclude zero indicate bias).

QuarkRMR® - Deltatrac®
Deltatrac® 2 – Deltatrac® 1a
95% confidence interval 95% confidence interval
Parameter (actual value) (as percentage of Deltatrac®) (actual value) (as percentage of Deltatrac® 1)
VO2 (mL/min)b −4.6 to 2.6 −1.8 to 1.0 −6.8 to 0.8 −2.7 to 0.3
VCO2 (mL/min)b −1.7 to 5.7 −0.8 to 2.7 −6.1 to 7.3 −2.9 to 3.8
RMR (kcal/day)b −27 to 22 −1.5 to 1.2 −43 to 10 −2.5 to 0.6
RQb −0.0005 to 0.0233 −1.8 to 1.0 −0.15 to 0.04 −1.7 to 4.6
a

Deltatrac® 2 = the second Deltatrac® measurement in Phase 1 and 3 when the Deltatrac® was used twice in the same measurement session. Deltatrac® 1 was the first measurement of the pair.

b

VO2 oxygen consumption, VCO2 carbon dioxide production, RMR resting metabolic rate, RQ respiratory quotient