Skip to main content
. 2014 Jul 15;472(10):2913–2922. doi: 10.1007/s11999-014-3784-6

Table 5.

Coronal, sagittal, and axial alignment differences between MRI and CT

Variable Deviation of variable from plan* p value
CT-based patient-specific instrumentation group MRI-based patient-specific instrumentation group
Long-leg mechanical axis 3.0° (0°–5°) 1.0° (0°–4°) 0.03
Mechanical axis of femur 1.0° (0°–2.5°) 1.0° (0°–2.5°) 0.18
Mechanical axis of tibia 1.0° (0°–2°) 1.0° (0°–1.5°) 0.45
Rotational alignment of femur 1.0° (0°–2°) 1.0° (0°–2°) 0.56
Posterior tibial slope (from 3°) 1.0° (0°–2°) 1.0° (0°–2°) 0.60

* Values are expressed as median, with range in parentheses.