Skip to main content
. 2014 Aug 24;14:74. doi: 10.1186/1471-2253-14-74

Table 3.

Statistical comparison of force values for 5 cm tube displacement or failure of fixator

Comparison of force values at 5 cm displacement or failure (N)
  SolidAIRity® Thomas™ Anchor Fast Tape & Malllinckrodt Twill & Mallinckrodt Tape & Securisyn Twill & Securisyn  
SolidAIRity®
 
0.000
0.000
0.000
0.000
0.000
0.007
P values
Thomas™
129 [109.13; 148.41]
 
0.065
0.530
0.651
0.036
0.000
Anchor Fast
107 [82.07; 132.35]
-22 [-44.59; 1.47]
 
0.371
0.032
0.645
0.000
Tape & Malllinckrodt
121 [92.35; 149.00]
-8 [-34.66; 18.46]
13 [-17.01; 43.93]
 
0.365
0.214
0.000
Twill & Mallinckrodt
132 [113.09; 151.53]
4 [-12.39; 19.46]
25 [2.40; 47.49]
12 [-14.66; 37.92]
 
0.019
0.000
Tape & Securisyn
100 [72.22; 128.52]
-28 [-54.78; -2.02]
-7 [-37.17; 23.48]
-20 [-53.13; 12.53]
-32 [-58.04; -5.83]
 
0.002
Twill & Securisyn
42 [12.60; 71.35]
-87 [-114.51; -59.08]
-65 [-96.67; -33.81]
-79 [-112.52; -44.87]
-90 [-117.79; -62.88]
-58 [-92.10; -24.70]
 
 
 
Difference between means and 95% confidence interval
 
Difference in standard deviation of means at 5 cm displacement or failure
 
SolidAIRity ®
Thomas™
Anchor Fast
Tape & Malllinckrodt
Twill & Mallinckrodt
Tape & Securisyn
Twill & Securisyn
 
SolidAIRity ®
 
0.723
0.311
0.795
0.267
0.002
0.008
P values
Thomas™
0.93 [-4.40; 6.26]
 
0.232
0.969
0.181
0.001
0.006
Anchor Fast
-4.26 [-12.90; 4.36]
-5.19 [-14.00; 3.62]
 
0.273
0.849
0.128
0.140
Tape & Malllinckrodt
0.80 [-5.54; 7.14]
-0.13 [-6.74; 6.49]
5.07 [-4.29; 14.42]
 
0.246
0.003
0.008
Twill & Mallinckrodt
-3.40 [-9.62; 2.81]
-4.33 [-10.83; 2.17]
0.86 [-8.42; 10.14]
-4.21 [-11.5; 3.09]
 
0.035
0.060
Tape & Securisyn
-11.49 [-18.05; -4.92]
-12.42 [-19.25; -5.59]
-7.22 [-16.71; 2.26]
-12.29 [-19.87; -4.71]
-8.08 [-15.56; -0.60]
 
0.857
Twill & Securisyn
-12.32 [-21.00; -3.65]
-13.5 [-22.10; -4.40]
-8.06 [-18.94; 2.82]
-13.12 [-22.51; -3.73]
-8.92 [-18.24; 0.40]
-0.83 [-10.35; 8.69]
 
 
  Difference between means and 95% confidence interval  

Statistical analysis of force values averaged across all test points for each fixation technique at 5 cm measured displacement. Confidence intervals were calculated using the Satterthwaite’s approximation (Welch’s t-test) due to likelihood of unequal variance. Difference values are the result of subtracting the mean value for the method in the row from the mean value for the method in the column (column mean-row mean). Differences showing statistical significance (α < 0.05) are listed in bold.