Table 3.

Comparison of ONH measurements generated by the merchant-provided software of the OCT machine and our algorithm regarding area (mm2) of disc, rim and cup and C/D ratio.

Parameter Normal Glaucoma
OCT* Algorithm* r p OCT* Algorithm* r p
Disc area (mm2) 2.50 ± 0.54 2.41 ± 0.39 0.88 0 2.97 ± 0.88 2.94 ± 0.86 0.86 4e-13
Rim area (mm2) 1.71 ± 0.69 1.64 ± 0.67 0.86 2e-12 1.45 ± 0.75 1.46 ± 0.78 0.82 7e-8
Cup area (mm2) 0.78 ± 0.51 0.77 ± 0.63 0.85 3e-19 1.52 ± 0.82 1.48 ± 0.79 0.80 9e-5
C/D Ratio 0.31 ± 0.14 0.32 ± 0.23 0.85 5e-17 0.51 ± 0.20 0.50 ± 0.34 0.81 3e-6
*

Mean ± STD;

Pearson correlation