Published in final edited form as: IEEE Trans Nucl Sci. 2014 Jun;61(3):1092–1098. doi: 10.1109/tns.2014.2302580

Table IV.

Average energy and timing resolution for the 56 crystals under the two scenarios discussed above. All measurements were performed with a 511 keV source and sufficient statistics. While we present average energy and timing resolution, measurement for each individual crystal had 5% precision.

8×7 array of diffuse sheet-reflector wrapped 1.5 × 1.5 × 12 mm3 LYSO crystals readout by the H8500 multi-anode PMT Using information from all the 64 (8 × 8) anodes By performing a software reduction of the 64 anodes to a 8+8 sum along the two orthogonal directions
Energy resolution (% FWHM) 13.8 ± 1.8 14.1 ± 1.9
Timing resolution (FWHM) 319 ± 11 ps 321 ± 10 ps