Table 2 Number of correctly detected simulated lesions using the digital mammography system and the detection rate in per cent averaged over the five evaluators, each three phantom configurations (c-c and ml plane [2D] and c-c and c-c plus 4° plane, respectively [3D]).

Amuletc-c and oblique plane2D Amuletc-c and c-c plus 4° plane3D Really existing lesions
Threads 18.0 (100 %) 18.0 (100 %) 18
Microcalcifications 14.3 (95.6 %) 14.3 (95.6 %) 15
Tumour-like masses 12.0 (100 %) 12.0 (100 %) 12
All lesions 44.4 (98.7 %) 44.4 (98.7 %) 45