Table 1.
Step | Problem | Possible cause | Solution |
---|---|---|---|
23 | Poor MALDI profiles (weak or no signals) during in vitro assay |
Reaction was not quenched properly |
Pipette the sample 10 to 20 times (Box 1 Step 2) to ensure proper mixing. Alternatively, vortex the samples after addition of 8 M GndHCl quenching solution. |
34 | Noisy quantitative data |
Reaction times were not carefully controlled |
Ensure that the reaction start and stop times are synchronized (Step 34) |