Published in final edited form as: Nat Protoc. 2013 Aug 15;8(9):1730–1742. doi: 10.1038/nprot.2013.104

Table 1.

Troubleshooting table.

Step Problem Possible cause Solution
23 Poor MALDI
profiles (weak or no
signals) during in
vitro assay
Reaction was not
quenched properly
Pipette the sample 10 to 20
times (Box 1 Step 2) to
ensure proper mixing.
Alternatively, vortex the
samples after addition of 8 M
GndHCl quenching solution.
34 Noisy quantitative
data
Reaction times
were not carefully
controlled
Ensure that the reaction start
and stop times are
synchronized (Step 34)