Skip to main content
. 2014 Sep 3;87(1042):20130654. doi: 10.1259/bjr.20130654

Table 7.

Effect of a reduced field of view (FOV) on effective dose (maxillofacial to full mandible)

Reference FOV1 (cm) FOV2 (cm) Dose reduction (%)a
Current study 17 × 12 14 × 5 32
Davies et al5 16 × 13 16 × 6 26
Roberts et al12 16 × 13 16 × 6 33
Schilling and Geibel15 16 × 13 16 × 4 67
Pauwels et al2 16 × 13 16 × 6 46
Morant et al16 16 × 13 16 × 6 42
a

Relative reduction of the tube current exposure time product-normalized effective dose of FOV2 compared with FOV1.