Table 2. Inter-assay precision of endogenous PCSK9 serum controls.


Serum Matrix
Controls
Mean ng/mL
(STD)
Inter-assay Precision
(%CV)
Precision of Individual Replicates (%CV) Acceptance Ranges (ng/mL)
SC 1 169
(18.7)
11 0 to 9 132 to 206
SC 2 411
(45)
11 0 to 7 321 to 501

Acceptance range: (mean ng/mL concentration ± 2STD)