Skip to main content
. Author manuscript; available in PMC: 2014 Oct 1.
Published in final edited form as: Nat Protoc. 2014 Mar 13;9(4):842–850. doi: 10.1038/nprot.2014.048

TABLE 1.

Troubleshooting table.

Step Problem Possible reason Solution
14 Low number of spectra Sample of high complexity (samples’ co-elution) Further fractionate or simplify the sample (i.e., by adjusting the HPLC gradient to make it more gradual to better separate peptides) and thus increase the number of spectra observed
Insufficient amount of injection or sample Determine the optimal amount of the sample for the type of MS instrument and LC gradient used. Adjust the amount of the sample loaded. Check the autosampler for possible failure
26 Low-quality tandem mass spectral data sets Problems with the chromatography column or system Use a simple protein digest (such as cytochrome c digest) to verify the integrity of the HPLC system
Insufficient amount of injection or sample Determine the optimal amount of the sample for the type of MS instrument and LC gradient used. Adjust the amount of the sample loaded. Check the autosampler for possible failure
Heavy contamination, such as polymers leaching from laboratory plasticware Use MS-compatible grades of plastic tubes, tips and powder-free gloves as indicated throughout the protocol
Low ‘spectrum-spectrum match ratio’ Low-quality tandem mass spectral data sets Open the raw data file and examine the total ion chromatogram by using the mass spectrometer’s software to check the quality of the data set. A good-quality data set should have numerous peaks that rise above the noise in the total ion chromatogram. If the quality is not good, see the troubleshooting section for low-quality tandem mass spectral data sets
The species of the query sample is not represented in the library Extend the library data set by adding more species