14 |
Low number of spectra |
Sample of high complexity (samples’ co-elution) |
Further fractionate or simplify the sample (i.e., by adjusting the HPLC gradient to make it more gradual to better separate peptides) and thus increase the number of spectra observed |
|
|
Insufficient amount of injection or sample |
Determine the optimal amount of the sample for the type of MS instrument and LC gradient used. Adjust the amount of the sample loaded. Check the autosampler for possible failure |
26 |
Low-quality tandem mass spectral data sets |
Problems with the chromatography column or system |
Use a simple protein digest (such as cytochrome c digest) to verify the integrity of the HPLC system |
|
|
Insufficient amount of injection or sample |
Determine the optimal amount of the sample for the type of MS instrument and LC gradient used. Adjust the amount of the sample loaded. Check the autosampler for possible failure |
|
|
Heavy contamination, such as polymers leaching from laboratory plasticware |
Use MS-compatible grades of plastic tubes, tips and powder-free gloves as indicated throughout the protocol |
|
Low ‘spectrum-spectrum match ratio’ |
Low-quality tandem mass spectral data sets |
Open the raw data file and examine the total ion chromatogram by using the mass spectrometer’s software to check the quality of the data set. A good-quality data set should have numerous peaks that rise above the noise in the total ion chromatogram. If the quality is not good, see the troubleshooting section for low-quality tandem mass spectral data sets |
|
|
The species of the query sample is not represented in the library |
Extend the library data set by adding more species |