12 |
Air bubble trapped in CTC-iChip1 lane |
Bubbles in buffer |
Clamp CTC-iChip1-to-CTC-iChip2 tubing, increase pressure to 140 kPa (20 psi) and tap on CTC-iChip1 to flush out bubbles via waste outlet |
14 |
Air bubble trapped in CTC-iChip2 channel |
Bubbles in buffer |
Unclamp CTC-iChip1-to-CTC-iChip2 tubing, press gently to guide the bubbles out from the outlets |
34 |
Blood level rises or blood not flowing through the system |
Pressure problem |
Make sure air pressure # 2 is on |
35 |
RBC carryover to CTC-iChip2 |
Pressure or syringe pump problem |
Check pressure and syringe pump settings, confirm that there is no leak |
35 |
Majority of CTC-iChip1 lanes blocked |
Blood has large aggregates |
Terminate the run |
37 |
Bubbles get into CTC-iChip1 |
Blood sample is finished, blood inlet not clamped on time |
Stop the run |