Table 1.

Troubleshooting

Step Problem Possible Reason Solution
12 Air bubble trapped in CTC-iChip1 lane Bubbles in buffer Clamp CTC-iChip1-to-CTC-iChip2 tubing, increase pressure to 140 kPa (20 psi) and tap on CTC-iChip1 to flush out bubbles via waste outlet
14 Air bubble trapped in CTC-iChip2 channel Bubbles in buffer Unclamp CTC-iChip1-to-CTC-iChip2 tubing, press gently to guide the bubbles out from the outlets
34 Blood level rises or blood not flowing through the system Pressure problem Make sure air pressure # 2 is on
35 RBC carryover to CTC-iChip2 Pressure or syringe pump problem Check pressure and syringe pump settings, confirm that there is no leak
35 Majority of CTC-iChip1 lanes blocked Blood has large aggregates Terminate the run
37 Bubbles get into CTC-iChip1 Blood sample is finished, blood inlet not clamped on time Stop the run