Table 2.
Summary of technical performance for 3 biomarkers assays
| C-Col10 (pg/ml) | hsCRP (μg/ml) | C2M (pg/ml) | |
|---|---|---|---|
| Assay specifications | |||
| Slope of standard curve | 1.23 | n. a. | 0.96 |
| IC50, pg/ml | 363 | n. a. | 440 |
| Intra-assay CV% | 4.19 | n. a. | 3.95 |
| Inter-assay CV% | 13.18 | <5 | 9.88 |
| Lower limit of detection | 24 | n. a. | n. a. |
| Quantifiable range | 40-7300 | 0.2-10 | 30-1700 |
C-Col10: C-terminus of ColX assay; C2M: MMP-derived collagen type II fragment assay; hsCRP: high sensitive CRP assay. n.a.: not available.