Table 1.
Analytical conditions for VOC with thermodesorption GC/MSD.
Thermal desorber UNITY/ULTRA (Markes, UK) |
GC/MSD (HP6890/5973, Hewlett Packard, USA) |
||
---|---|---|---|
Oven temp. | 300°C | GC column | Rtx-1 (0.32 mm, 105 m, 1.5 μm) |
Desorb time | 10 min | Initial temp. | 50°C (10 min) |
Desorb flow | 50 mL/min | Oven ramp rate | 5°C/min |
Cold trap holding time | 5 min | Final temp. | 250°C (5 min) |
Cold trap high temp. | 320°C | Post run | 250°C (5 min) |
Cold trap low temp. | −10°C | Column flow | 1.13 mL/min |
Cold trap packing | Tenax TA/carbopack B | Detector type | Quadropole |
Min. pressure | 12 psi | Q-pole temp. | 150°C |
Inlet split | No | MS Source temp. | 230°C |
Outlet split | 10 mL/min | Mass range | 35~300 amu |
Valve and line temp. | 180°C | Electron energy | 70 eV |