Table 1.

Intravascular and angiographic imaging analyses at baseline and at 28-day follow-up following PCB delivery in porcine coronary arteries

Crystalline coating (n=8) Amorphous coating (n=9) POBA control (n=9)
QVA
Baseline
 RVD (mm) 3.09±0.1 3.00±0.1 3.05±0.1
 Stent to artery ratio 1.46±0.1 1.35±0.1 1.48±0.1
 Balloon to artery ratio 1.46±0.1 1.41±0.1 1.46±0.1
 Poststent MLD (mm) 3.28±0.2 3.08±0.3 3.13±0.2
OCT
28-day follow-up
 RVD (mm) 2.69±0.3 2.65±0.2 2.60±0.2
 MLD (mm) 1.62±0.6 1.41±0.3 1.25±0.7
 Diameter stenosis (%) 38.8±24 46.9±10. 51.8±25
 Lumen area (mm2) 5.08±0.8* 2.76±0.28 3.1±2.3
 Stent area (mm2) 10.65±0.4 9.12±0.3† 10.45±1
 Neointimal area (mm2) 5.57±0.7 6.36±0.3 7.20±1.8
 Area stenosis (%) 53.1±14* 69.7±11 71.2±18
 Neointimal thickness (mm) 0.63±0.1 0.80±0.1 0.80±0.2

*p<0.05 vs Amorphous and POBA.

†p<0.05 vs Crystalline and POBA.

MLD, minimal lumen diameter; OCT, optical coherence tomography; PCB, paclitaxel-coated balloons; POBA, plain balloon angioplasty; QVA, quantitative vessel analysis; RVD, reference vessel diameter.