Skip to main content
. 2014 Sep 30;13:140. doi: 10.1186/1475-925X-13-140

Table 2.

CURE Soft application procedure

Online mode Offline mode
Ventilator settings
1. Check ventilator output, network communication, output waveform is selected with baud rate 38400. -
2. Connect an RS232 serial port from the rear of the ventilator to the computing platform with CURE Soft. -
Computer settings
1. Execute CURE Jar. 1. Execute CURE jar.
2. In settings tab, select serial port 2. In settings tab, select captured file.
3. Select the serial port connected to the RS232. 3. Select the designated captured file (*.txt).
4. Click ‘start collecting data’. 4. Click ‘start collecting data’.