Table 2.

Treatment parameters (chronological order)

Treatment Duration (min) N treatment cells ( n sonications) Treatment cell diameter (mm) Acoustic power (W) median, range Sonication energy (kJ) median, range Total delivered energy (kJ) Temperature (°C) median, range Ablation approach b
1aa
40
2 (10)
4
100 (40–140)
2.0 (0.8–2.8)
17.5
-
Soft tissue
1ba
20
3 (6)
2
50 (40–60)
1.0 (0.8–1.2)
6.0
-
Soft tissue
2
41
5 (6)
2, 4
80 (50–100)
1.5 (0.9–1.9)
8.8
43.9 (39.3–49.0)
Soft tissue
3
37
6 (10)
4
110 (90–120)
1.8 (1.1–1.9)
16.6
63.1 (56.3–74.6)
Near field
4
45
7 (15)
4, 8
120 (60–160)
2.2 (1.0–3.0)
31.3
47.3 (43.1–50.2)
Soft tissue
5a
37
10 (11)
4
50 (50–70)
0.8 (0.8–1.1)
10.2
56.2 (48.1–60.7)
Near field
6
73
18 (28)
4
100 (50–130)
1.6 (0.6–2.1)
42.4
55.2 (45.3–60.4)
Near field
7
49
16 (19)
4
130 (70–150)
2.1 (1.1–3.0)
39.1
64.8 (51.3–75.5)
In front of cortexc
5b
69
20 (22)
4,8
40 (30–50)
0.6 (0.4–1.0)
14.9
60.9 (48.4–70.9)
Direct
8
68
22 (25)
4
120 (90–150)
1.9 (1.4–2.4)
48.4
56.1 (40.5–65.6)
Near field
9
71
21 (22)
8, 12
150 (70–150)
3.0 (2.0–3.0)
62.6
51.4 (41.9–67.8)
Soft tissue
10
39
9 (15)
4
20 (20–20)
0.3 (0.3–0.3)
4.2
63.0 (52.3–67.5)
Direct
11 45 12 (12) 4 40 (40–60) 0.6 (0.6–1.0) 7.4 64.4 (58.4–74.8) Direct

Abbreviations: °C degrees Celsius, mm millimeter, min minutes, kJ kilojoule, W watt, CR complete response, PR partial response, NR no response.

aIn treatment 1a and 1b, no representative temperature data were available due to susceptibility artifacts caused by internal fixation material. Treatment 1b was terminated early due to severe susceptibility artifacts (targeted volume too close to fixation material).

bDirect approach: the treatment cell was positioned on the cortical surface. Near-field approach: the treatment cells were positioned 10 mm distal to the cortical bone. Soft-tissue approach: in lesions with complete cortical destruction, the treatment cells were positioned in the tumor mass itself.

cBecause of lungs in the far field, for safety reasons, treatment cells were positioned in the tumor mass in front of the cortical bone.