Table 2.

Emergency and transport ventilators’ general characteristics (simple and mass-casualty/military)

Device Company Technology Pediatric use (≥10 kg) FIO 2 (range/%) Circuit Built-in PEEP Weight (kg) Volume (dm 3 ) Screen (with curves) Sonometry function (min/max, dB) Sonometry alarm (max, dB) Oxygen consumption FIO 2 100%/50% or Air-Mix (L/min) Maximal autonomy FIO 2 100% cylinder E (min) Electrical autonomy (min) Market price ($ US)
Simple emergency and transport ventilators (n = 9)
Autovent 3000 Allied Healthcare Products (Saint Louis, MO, USA) Pneumatic Yes 100% only Mono No 2.8 4.8 No 47.2/48.0 55.3 9.7/NA 70 NA 5,000 to 10,000 US $
Carevent MRI O-Two (Mississauga, ON, Canada) Pneumatic Yes 100%/Air-Mix Mono Yes 2.5 5.4 No 43.2/52.7 78.7 10.0/5.2 68 NA 5,000 to 10,000 US $
M. Standard Weinmann (Hamburg, Germany) Pneumatic Yes 100%/Air-Mix Mono No 1.2 4.0 No 42.9/44.3 49.8 8.5/3.5 80 NA 5,000 to 10,000 US $
Osiris 2 Air Liquide Medical Systems (Antony, France) Pneumatic Yes 100%/Air-Mix Mono Yes 5.0 9.9 No 47.3/53.3 83.0 11.4/5.3 56 NA 5,000 to 10,000 US $
Oxylog 2000 Dräger (Lübeck, Germany) Pneumatic Yes 100%/Air-Mix Mono No 4.3 5.3 No 46.6/61.6 78.8 9.4/4.7 72 ND 5,000 to 10,000 US $
Oxylog 2000 + Dräger Pneumatic Yes 100%/Air-Mix Mono Yes 5.4 8.4 Yes 46.2/52.2 77.0 7.7/2.1 88 ND 10,000 to 15,000 US $
ParaPAC 210 D Smiths Medical (Ashford, UK) Pneumatic Yes 100%/Air-Mix Mono No 3.2 3.0 No 43.9/45.6 56.6 8.8 /3.3 77 NA 2,500 to 5,000 US $
ParaPAC 310 Smiths Medical (Ashford, UK) Pneumatic Yes 100%/Air-Mix Mono Yes 2.3 3.6 No ND ND ND ND NA 2,500 to 5,000 US $
pNeuton Airon (Melbourne, FL, USA) Pneumatic No 100%/Air-Mix Mono Yes 3.5 7.9 No 43.4/61.3 84.9 11.1/NA 61 NA 5,000 to 10,000 US $
Mass-casualty and military ventilators (n = 2)
ComPAC Smiths Medical (Ashford, UK) Pneumatic/compressor No 100% O2/Air-Mix or 100% air Mono No 11.1 18.0 No 41.7/65.8 42.0 7.1/2.4 95 78 5,000 to 10,000 US $
MCV100 Allied Healthcare Products (Saint Louis, MO, USA) Pneumatic/compressor No Only 100% O2 or 100% air Mono No 6.9 9.0 No 43.0/48.3 61.8 7.1/NA 96 298/1,142 5,000 to 10,000 US $

Pediatric use, availability of a pediatric mode; FIO2, oxygen inspiratory fraction (%); Circuit, type of circuit to be used, monobranched, dual limb, both types; PEEP, positive end expiratory pressure; Volume, the volume of the device was calculated as assimilated to a cube; Sonometry, provides the noise level in function - values are provided without added PEEP and for FIO2 = 100% for most values, for FIO2 = 100%/50% for the two other; Autonomy was evaluated at standardized settings - MCV 100 electrical autonomy has been calculated while using the compressor; Market price has been considered taking into account mean value provided by various distributors on the Canadian and the European market. The air-mixing condition varies from one ventilator to the other (FIO2 = 45 to 86%), but also for the same ventilator, according to patient’s condition (FIO2 AirMix Osiris = 69 to 81%; O2000 = 68 to 80%); it may induce up to a 30% peak-flow decrease with several devices; this flow decrease at a constant I/E ratio is responsible for a tidal volume decrease (FIO2 or Venturi effect). NA= not applicable; ND= not done.