Skip to main content
. Author manuscript; available in PMC: 2014 Oct 15.
Published in final edited form as: Acad Radiol. 2011 Mar 25;18(6):705–715. doi: 10.1016/j.acra.2011.01.006

Table 2.

Scanning parameters for T1W, T2W, SENC and different variations of FSENC scans. T1W and T2W are scanned while the device is in compress and hold position.

Scan In-plane resolution (mm) Slice thickness (mm) W0 =W0 mm−1 WH mm−1 TFE
TSE
EPI Compression cycle/slice Total Scan time (sec) for 5 slices
T1W 1×1 5 - - TSE=3 - - 39
T2W 1×1 5 - - TSE=12 - - 67
SENC 1×1 5 1.132 1.332 TFE=10 Non 19 190
SENC1 2×2 5 1.132 1.332 TFE=10 Non 9 90
SENC2 3×3 5 1.132 1.332 TFE=10 Non 6 60
SENC3 3×3 10 0.567 0.667 TFE=10 Non 6 60
FSENC1 2×2 5 1.132 1.332 TFE=17 3 1 10
FSENC2 3×3 5 1.132 1.332 TFE=11 3 1 10
FSENC3 3×3 10 0.567 0.667 TFE=11 3 1 10