Table 2. Accuracy and precision of the method for detection of SP-141 in mouse plasma samples.

Nominal Concentration
(ng/mL)
Intra-Day (n=5)
Inter-Day (n=4)
Measured
(ng/mL)
Precision
(%CV)
Accuracy
(%RE)
Measured
(ng/mL)
Precision
(%CV)
Accuracy
(%RE)
LLOQ (0.648) 0.645±0.010 1.6 −0.5 0.611±0.025 4.2 −5.6
LQC (1.94) 1.89±0.02 1.3 −2.6 1.82±0.07 3.9 −6.0
MQC (81.0) 76.0±1.2 1.6 −6.1 78.3±2.4 3.1 −3.3
HQC (129.6) 123.9±0.6 0.5 −4.4 125.8±1.8 1.4 −3.0
UHQC1 (3240.0: 25-fold) 3310.1±11.8 0.4 2.1 3215.0±22.4 0.8 −0.9
UHQC2 (6480.0: 50-fold) 6595.1±35.8 0.6 1.3 6607.5±47.5 1.2 0.8