Skip to main content
. 2014 Aug 19;86(20):10036–10043. doi: 10.1021/ac5014767

Table 1. Linearity, Detection Limit (DL) S/N, and DL Mass Error for SIM Response Curves in Figure 4C and DL Data in Supplemental Figure S4 in the Supporting Information.

analyte slope adjusted R2 % RSD detection limit (pg) DL S/Na DL mass error (ppm)b
phenol-d6 0.9244 0.9986 24.7 0.005 5.89 ± 0.55 2.02 ± 2.78
hexachloroethane 0.9262 0.9996 23.9 0.010 5.02 ± 2.20 1.73 ± 2.64
2,4-dichlorophenol 0.8719 0.9911 47.9 0.005 10.2 ± 0.83 3.09 ± 0.57
      18.2 0.025 24.0 ± 3.85 4.36 ± 0.74
hexachlorobutadiene 0.9448 0.9985 28.5 0.005 9.22 ± 0.90 2.68 ± 0.78
2,4-dinitrotoluene 1.0443 0.9950 32.6 0.010 3.98 ± 2.46 6.65 ± 0.91
a

Average detection limit spectral signal-to-noise ratio (±standard deviation) at peak apex.

b

Average detection limit mass error (±standard deviation) at peak apex.