Table 1.

Global RMSEs of the reconstructed PRF images of different time frames from the 12- frame time series. For TCR and Complex Difference, RMSEs for reconstructions using only current time frame (denoted as “Single”), current and previous time frames (denoted as “Previous”), and using all time frames (denoted as “All”) are reported.

Time Frame R=1 (1 Average) R=3, Low Resolution R=3, Separated CS R=3, TCR R=3, Complex Difference
Previous All Single Previous All
#2 43% 69% 92% 83% 93% 57% 57% 57%
#3 28% 47% 56% 61% 64% 41% 40% 40%
#4 21% 33% 42% 45% 48% 32% 30% 29%
#5 16% 30% 35% 38% 41% 26% 25% 23%
#6 13% 26% 27% 35% 35% 23% 22% 21%
#7 11% 26% 26% 33% 31% 21% 20% 19%
#8 10% 26% 23% 29% 28% 20% 18% 17%
#9 8.7% 24% 23% 26% 26% 20% 17% 16%
#10 7.5% 24% 23% 25% 23% 17% 16% 15%
#11 6.6% 24% 20% 23% 23% 18% 15% 14%
#12 6.5% 24% 19% 22% 22% 17% 14% 14%