Table 2.

The plan conformity indexes, all fitting parameters from Equation (1), and the total number of monitor units (MUs) for the stereotactic radiosurgery plans with 90–50% prescription isodose levels (IDLs) for an example patient (Patient 2)

IDL 90% 80% 70% 60% 50%
CI 1.230 1.230 1.230 1.221 1.230
a1 15.720 11.000 10.181 9.940 9.760
b1 0.136 0.239 0.280 0.276 0.260
a2 3.506 8.248 9.032 9.437 9.315
b2 0.0371 0.0620 0.0651 0.0648 0.0634
Total MU 2465 2792 3200 3755 4531

CI, conformity index.

For this patient, the plan with 70% IDL achieved both the greatest dose-dropping speed (b1), and the greatest b2, the smaller exponential decay coefficient in equation (1) characterizing the dose drop-off in the low dose region.