Skip to main content
. Author manuscript; available in PMC: 2014 Oct 23.
Published in final edited form as: Lab Chip. 2013 May 7;13(9):1790–1796. doi: 10.1039/c3lc41269e

Table 1.

Flow rates for inlet and outlets A-E, plasma yield and the amount of cells/L of the generated plasma for the improved acoustic plasmapheresis microchip as compared to our previously reported microchip design.

Inlet (μl/min) Waste outlet A (μl/min) Waste outlet B (μl/min) Waste outlet C (μmin) Waste outlet D (μl/min) Plasma outlet E (μl/min) Fraction of plasma from whole blood Erythrocytes/L in produced plasma
Previous design 80 20 20 15 15 10 12.5 % 3.7×109
Improved design 50 10 10 10 10 10 20% 4.2×109