Skip to main content
. Author manuscript; available in PMC: 2014 Nov 1.
Published in final edited form as: Anal Bioanal Chem. 2014 Sep 6;406(28):7117–7125. doi: 10.1007/s00216-014-8058-3

Fig. 2.

Fig. 2

a Sample recoveries after depletion of human plasma with micro-scale online depletion as a function of sample introduction flow rate. FF yield (left axis) showed a linear increase with flow rate (R2=0.93). BF yield did not show significant correlation with flow rate (R2=0.5). The reported values are the mean of two technical replicates. FF flow-through fraction, BF bound fraction. The overall standard deviations of four flow rates were 5.1 % (FFs) and 4.2 % (BFs), respectively. b Measured depletion efficiency of microscale column at different flow rates compared with depletion efficiency of commercial IgY14 column (commercial column) operated at the manufacturer suggested flow rate