Figure 4.

Figure 4

The 62Cu elution yield on the day of generator receipt plotted as a function of the elapsed time since the preceding elution for 42 generators. The elution yield was observed to decline, and become more variable, when the elapsed time between elutions was greater than 60 minutes.