Table 3.
Overall determination of the accuracy of the GlucoRx Nexus TD-4280 blood glucose monitoring device relative to the accuracy limits defined in both the ISO 15197:2003 and ISO 15197:2013 guidelines.
| DIN EN ISO 15197:2003 | DIN EN ISO 15197:2013 | |
|---|---|---|
| Accuracy limits set by guideline | ±15 mg/dL (0.83 mmol/L) and ±20% | ±15 mg/dL (0.83 mmol/L) and ±15% |
| Percentage of results required to be within this limit for meter to be deemed accurate | ≥95% | ≥95% |
| Number of results obtained within this limit | 97/105 | 91/105 |
| Percentage of results obtained within this limit (%) | 92.4 | 86.7 |
| Are ≥95% of the readings within the accuracy limits set in guidelines? (Yes/no) | No | No |