Published in final edited form as: Curr Med Imaging Rev. 2012;8(1):16–26. doi: 10.2174/157340512799220616

Table 3.

Summary of HMI heating rates and lesion sizes for 10- 20- and 30-s HIFU exposures in seven (n = 7) porcine liver specimens, i.e., 21 sonication locations total

Sonication
duration (s)
HMI heating rate before
lesion formation (µm/°C)
HMI heating rate after
lesion formation (µm/°C)
Lesion size

Diameter
(mm)
Length
(mm)
10 0.87 ± 0.082 −0.88 ± 0.080 2.0 ± 0.12 6.4 ± 1.14
20 0.85 ± 0.12 −0.71 ± 0.16 3.4 ± 0.13 13.1 ± 1.15
30 0.80 ± 0.14 −0.79 ± 0.19 4.7 ± 0.18 16.4 ± 0.39