Table II.

Imprecision (CV) and bias for 35 analytes. Results obtained after storage of whole blood at 21°C was compared to baseline results.

CV (%)
Bias (%)
Paired t-test p value††
Analyte CV Analytical/goal§ Mean deviation (95% CI)|| Goal
ALAT 2.4 2.7/9.0 − 1.7 (− 2.9;− 0.4) 12.0** 0.0034
ALB 1.5 0.9/3.2** 0.2 (0;0.5) 2.2** 0.067
AMYL 1.1 2.4/5.9 − 0.8 (− 1.2;− 0.3) 8.0 < 0.0001
ATRYP 2.9 0.9/4.5 − 0.7 (− 1.5;0.1) 6.5 0.12
B12* 6.4 6.4/7.5 − 3.2 (− 5.6;− 0.8) 17.7 0.015
BASP 1.2 2.2/4.8 0.2 (− 0.3;0.7) 6.4 0.31
BILI 2.1 3.2/6.4** 1.0 (0.2;1.9) 10.0** 0.11
CA 1.1 1.0/2.5** 1.0 (0.7;1.3) 2.0** < 0.0001
CHOL 1.3 0.5/3.0** 0.6 (0.2;1.1) 4.0 0.0085
CK 1.4 0.8/5.7 − 0.7 (− 1.3;− 0.1) 11.5 0.0023
CO2 3.4 5.0/6.0** 1.2 (0.0;2.4) 5.0** 0.16
CREA 1.1 1.3/2.7** − 0.1 (− 0.6;0.3) 3.8** 0.50
CRP 1.8 1.8/5.0** 2.6 (− 0.5;5.7) 5.0** 0.046
F-T3* 2.4 2.2/7.5** 0.8 (− 0.3;1.8) 7.2 0.28
F-T4* 4.6 4.6/7.5** 2.1 (0.3;3.9) 3.6** 0.039
Fe 3.7 1.6/6.7** 6.8 (5.8;8.2) 8.8 < 0.0001
FER* 7.0 4.3/7.1 − 1.1 (− 2.7;0.5) 5.2 0.039
FOL* 20.5 6.6/12.0 − 13.7 (− 19.1;− 8.3) 19.2 0.0007
GGT 2.6 1.6/3.5 − 0.8 (− 2.1;0.6) 5.4 0.62
HAPT 1.7 1.4/5.1 0.1 (− 0.4;0.6) 10.4 0.72
HDL-C 2.9 1.8/3.6 4.6 (3.7;5.4) 5.2 < 0.0001
IgA 2.1 1.3/2.7 1.9 (1.4;2.3) 9.1 < 0.0001
IgG 1.0 0.9/2.3 0.2 (− 0.2;0.5) 4.3 0.75
IgM 1.5 2.5/4.5** 1.7 (0.9;2.5) 11.9 0.0001
K 2.9 0.8/2.4 1.2 (0.1;2.2) 1.8 0.040
LD 4.5 3.1/5.0** 1.0 (− 1.3;3.1) 4.3 0.55
Na 0.4 0.7/1.0** 0.3 (0.1;0.4) 0.6** 0.0003
OROS 1.5 0.9/5.7 0.8 (0.1;1.5) 6.8 0.017
P 9.0 1.8/4.3 − 11.2 (− 12.4;− 10.0) 3.2 < 0.0001
PSA* 3.4 3.4/9.1 − 0.4 (− 2.0;1.2) 18.7 0.083
TGLY 2.0 1.0/5.3** − 0.9 (− 1.7;− 0.1) 5.4** 0.018
TRANS 1.5 0.9/2.3 1.8 (1.4;2.2) 2.6** < 0.0001
TSH* 9.2 4.9/9.7 6.0 (3.4;8.7) 7.8 0.0040
URATE 1.6 1.3/4.3** 1.5 (0.9;2.1) 4.8** < 0.0001
UREA 1.0 1.9/6.2 − 0.2 (− 0.6;0.2) 5.5 0.072
*

Analysis performed on Advia Centaur XP using whole blood in serum tubes. For the other analytes analysis was performed on Architect c8000 using whole blood in lithium-heparin tubes.

CV calculation based on measurements at baseline and after 10 h of storage.

Annual (2011) average CV of laboratory routine quality control samples.

§

CV goals used in evaluating analytical variation in laboratory routine quality control. Goals were generally based on biological variation from http://westgard.com/biodatabase1.htm and calculated as described by Fraser et al. [18].

**

denotes goals where empirically based modifications to the three stated levels of performance were used.

||

Mean deviation % between baseline and 10-h values and 95% confidence interval (shown as confidence limits).

Bias goals used in evaluating systematic differences in laboratory routine quality control.

See §CV goals for further details.

††

p-values for testing baseline values against values obtained after 10 h of storage of whole blood.