Table 1.

Technical specifications of the slit-based and coded aperture instruments. Notice that the coded aperture instrument provides a larger collection area than the slit-based instrument (15.4×106 versus 52.0  μm2) while maintaining a similar spectral resolution and collection numerical aperture (NA). This is because of the sharp difference in étendue, which is nearly 6 orders of magnitude greater for the coded aperture instrument.

  Slit based Coded aperture
Illumination wavelength (nm) 785 785
Illumination power (mW) 10 85
Illumination area (μm2) 80 (circular) 8×106 (elliptical)
Working distance (mm) 10.6 1.5
Spectral resolution (cm1) 10 7
Collection NA 0.5 0.45
Collection area (μm2) 52.0 (2×26) 15.4×106 (4400×3500)
Étendue (μm2sr) 43.8 10.4×106