Table 1.

Sensitivity/specificity (%) summary table using per lesion analysis. Accuracy is the percentage of correct classifications [(true positive+true negative)/total lesions]. For instance, for SCCBCC versus AK, true positive=0.95×71=54, true negative=0.71×14=10, total lesions=71, and accuracy=(54+10)/71=90%. Best diagnostic performance with the least number of modalities and/or PCs are in bold.

Classifier # Lesions Sensitivity/Specificity (%) Accuracy (%)
RS DOS LIFS Combined
MM versus PL 12 versus 17 100/100R3, R5, R8, R4, R9 17/59 67/18 100/100 100
MMPL versus norm. 29 versus 28 90/82 97/100 93/100 100/100(D1, R9) or (L1, R9) 100
SCCBCC versus AK 57 versus 14 72/64 75/71 91/57 95/71D2, L2, R9 (54+10)/71=90
AKSCCBCC versus norm. 71 versus 71 68/55 87/68 52/52 90/85D1, D2, L1, L2, R7 (64+60)/142=87