Skip to main content
. 2013 Nov 25;13:73. doi: 10.1186/1471-2415-13-73

Table 3.

The means and standard deviations for CCT measurements obtained using the SP-2000P and z-ring CS4, intra-observer repeatability, inter-observer reproducibility (z-ring CS4 only), and comparison between SP-2000P vs. z-ring CS4 (mean differences, limits of agreement, paired t -test, un-paired test, and p-values)

Analysis of CCT values
Observer 1
Observer 2
  USP SP-2000P z-ring CS4 z-ring CS4
Mean ± SD*
555.11 ± 35.83
535.82 ± 41.10
552.57 ± 36.83
554.97 ± 36.34
Intra-observer repeatability (p-value)
0.91
0.20
0.45
0.54
 
Mean difference ± SD*
Limits of agreement
p-value
Inter-observer reproducibility with z-ring CS4**
-2.40 ± 9.10
+39.90, -38.70
0.16
USP vs. SP-2000P
19.30 ± 19.60
+57.70, -19.20
 
USP vs. z-ring CS4
2.50 ± 5.70
+16.30, -11.20
SP-2000P vs. z-ring CS4 -16.70 ± 18.60 +23.70, -57.20  

Standard deviation*.

Paired test**.