Figure 2.

Figure 2

T1 maps from a healthy subject (subject #5), acquired using the proposed and uniform point selection strategies. The T1 map generated using the proposed point selection is more homogenous, as visible in the blood pool. prec(T1est) averaged over the 5 acquisitions were 105.1 ± 18.9 ms and 81.8 ± 10.3 ms for the myocardium as measured by reader 1, and 95.3 ± 6.8 ms and 76.6 ± 5.0 ms as measured by reader 2 for uniform and proposed selection strategies respectively. prec(T1est) for the blood pool were 150.2 ± 15.1 ms and 116.8 ± 9.6 ms as measured by reader 1, and 164.7 ± 18.9 ms and 114.2 ± 10.4 ms as observed by reader 2 for uniform and proposed selection strategies respectively.