Published in final edited form as: ASAIO J. 2014 Jul-Aug;60(4):391–399. doi: 10.1097/MAT.0000000000000093

Table 2.

Hemodynamic effects (mean ± SD) both with and without a valve.

Baseline No Valve Contegra valved conduit

MEAN±SD EXP PVC25 PVC50 EXP PVC25 PVC50
PTCPC [mm Hg] 14.1±1.3 14.2±1.8 14.2±1.9 14.2±1.3 14.5±1.8 14.7±2.0
PLiver 15.6±1.6 15.8±2.3 15.8±2.3 15.3±1.7 15.3±2.2 15.1±2.3

QIVC [Lpm] 2.25±1.6 2.26±2.0 2.29±2.2 2.25±1.5 2.23±1.8 2.26±1.8
QHV 0.50±1.0 0.51±1.1 0.53±1.2 0.50±0.9 0.50±1.0 0.52±1.0
Qsvc 1.14±0.3 1.15±0.5 1.15±0.5 1.15±0.4 1.17±0.5 1.18±0.5
CO 3.39 3.41 3.43 3.40 3.40 3.44

Qante IVC 136.2 143.1 148.2 133.4 133.2 136.6
Qretro IVC 4.34 10.0 13.3 2.0 2.8 4. 5
QretroIVCQanteIVC
[ml/tR] 3.2% 7.0% 9.0% 1.5% 2.1% 3.3%
Qante HV 36.4 37.3 44.4 35.8 35.4 39.8
Qretro HV 7.1 9.1 14.2 6.8 6.1 9.7
QretroHVQanteHV
20% 22% 32% 19% 17% 24%

Power Loss [mW] 24.4 26.0 28.2 21.8 22.3 24.7