Table 3. K2 and K3EDTA Plasma Equivalence and Dilution Accuracy.
DTG in Plasma Matrix | |||||||
---|---|---|---|---|---|---|---|
EDTA Equivalence | K2EDTA | K3EDTA | |||||
| |||||||
Low | Mid | High | Low | Mid | High | ||
| |||||||
15 | 450 | 9000 | theoretical conc. | 15 | 450 | 9000 | |
14.10 | 412.67 | 8676.90 | mean | 15.83 | 460.87 | 9236.90 | |
1.02 | 22.75 | 59.93 | SD | 0.31 | 20.72 | 149.29 | |
7.2 | 5.5 | 0.7 | %CV | 1.9 | 4.5 | 1.6 | |
-6.0 | -8.3 | -3.6 | %dev | 5.6 | 2.4 | 2.6 | |
-9.2 | -3.5 | -3.3 | %diff* | 9.0 | 3.5 | 3.3 | |
| |||||||
Dilutions of Ultra High QC | K2EDTA | K3EDTA | |||||
| |||||||
1/5 dil | 1/10 dil | 1/20 dil | 1/5 dil | 1/10 dil | 1/20 dil | ||
| |||||||
6000 | 3000 | 1500 | theoretical conc. | 6000 | 3000 | 1500 | |
5477.57 | 3004.63 | 1459.60 | mean | 5615.77 | 2863.67 | 1457.73 | |
173.29 | 24.99 | 17.13 | SD | 184.31 | 58.19 | 45.08 | |
3.2 | 0.8 | 1.2 | %CV | 3.3 | 2.0 | 3.1 | |
-8.7 | 0.2 | -2.7 | %dev | -6.4 | -4.5 | -2.8 |
%difference as compared to matrix matched QCs, determined by the following equation: (mean unmatched – mean matched)/[((mean unmatched + mean matched)/2)*100].